Ponemah Closing after performing attribute analysis on BPR channel
Issue: Ponemah suddenly closes after BPR analysis is performed.
SYMPTOMS:
When loading Ponemah in a review session the file will load as expected with all channels. Analysis attributes can be computed for all channels without any issues. When the user selects to analyze the BPR channel with the defined attribute settings, Ponemah will unexpectedly close. This will occur when there is a difference in sampling rate between the BP channel and the BPR channel.
SOLUTION:
The issue can be avoided by keeping the minimum Respiration Smoothing attribute 6 or larger.