Model: Communication Link Controller (CLC)
Version: 1.29.206
Date: August 2017
Firmware Release Notes for CLC version 1.29.206 indicate revisions made to the CLC Firmware since its initial release.
Compatibility
These firmware updates are compatible with all PhysioTel™ Digital Implants and hardware. Version 1.29.206 is compatible with Ponemah versions 5.10, 5.20, Ponemah 6.xx and all associated service packs. Ponemah 5.2 SP9 and 6.33 need to use CLC firmware 0.1.28, or 1.29.206.
These firmware updates are compatible with all PhysioTel™ Digital Implants and hardware. Version 0.1.28 IS NOT compatible with Ponemah versions 5.10, 5.20, and all associated service packs, as well as Ponemah 6.32 and previous versions of Ponemah 6.xx.
Release History
Version |
0.1.20* |
0.1.21 |
0.1.22/23 |
0.1.24 |
0.1.25 |
0.1.26 |
0.1.27 |
0.1.28 |
1.29.206 |
Release Date |
December 2012 |
January 2014 |
DSI Internal Releases |
April 2014 |
February 2016 |
June 2016 |
October 2016 |
August 2017 |
September 2017 |
*Original firmware released with CLC.
New features, enhancements, and fixes
Reference # |
Release Version |
Description |
Disposition |
37840 |
1.29.206 |
Improving process for reconnecting with implants that were out of range, and have come back into range. Will result in a quicker re-connection. |
Enhancement |
38676 |
1.29.206 |
Change the CLC firmware revision scheme to include a unique build number. (Releases changing from 0.1.28, to 1.29.xxx) |
Enhancement |
39292 |
1.29.206 |
CLC will now blink the yellow TRX LED if there is an error with the TRX, or if there is a CRC error between the communication of the CLC and TRX. |
Enhancement |
39401 |
1.29.206 |
CLC can get stuck in a loop if implant is out of range while doing scheduled sampling. It would not be possible for the CLC to every acquire data from that implant again even if it was brought back into range. |
Fix |
39514 |
1.29.206 |
Fixes a bug when using new CLC FW versions with the older (SP8, SP6) configuration or wizard process. It was not possible to proceed with the wizard if changing the CLC frequency. |
Fix |
39555 |
1.29.206 |
TRX Power can be adjusted by EU & CH Users |
Enhancement |
37813 |
0.1.28 |
Fix a bug within CLC to show the correct status of the implant in Ponemah 6.x |
Fix |
37512 |
0.1.28 |
Fix issue of the CLC reporting an impossibly high battery voltage to DSILink |
Fix |
36961 |
0.1.28 |
Improve the shutdown/reboot process of the CLC to prevent lockups |
Fix |
37023 |
0.1.28 |
CLC Logs can now be obtained by a button on the System Status page of the CLC Webpage |
Enhancement |
38303,38305 |
0.1.28 |
Decreasing amount of data written to SD card, and adding ability to stop writing data to SD card by checkbox on CLC Settings Page |
Fix |
37941,37940 |
0.1.28 |
Fixes a data loss bug within the CLC dealing with reading/writing data from memory |
Fix |
36810 |
0.1.28 |
CLC will no longer support Listen Before Talk in EU/Japan while in Assessement mode. This will prevent the TRX reporting an error of an unrecognizable command |
Fix |
36738 |
0.1.28 |
Update to support the Frame Beacon Lock to prevent implants from listening to beacons from adjacent systems |
Enhancement |
36812 |
0.1.28 |
CLC will notify DSILink of an implant reset as a reset instead of an error. CLC will also record what caused the reset in its logs, and send to DSILink as well |
Fix |
35660 |
0.1.28 |
TRX Power can be adjusted by EU only Users |
Enhancement |
34885 |
0.1.28 |
PhysioTel Digital China support |
Enhancement |
34887,34888,35279, 35156,38306,38322 |
0.1.28 |
Updates to support 6 implants per CLC |
Enhancement |
386677, 38381, 37195 |
0.1.28 |
Updates to improve user experience with DSILink 1.9 |
Enhancement |
34470 |
0.1.28 |
Fix for uploading TRX firmware |
Fix |
35270 |
0.1.27 |
Fix for implant channel signal misalignment when sampled using different sampling rates. |
Fix |
34804 |
0.1.26 |
Fixed an issue that caused Ponemah v5.xx and v6.00 to crash when launching the PhysioTel Digital configuration if a CLC with firmware 0.1.25 was on the same network as the acquisition computer. |
Fix |
34355, 34470, 33515 |
0.1.26 |
Fixes to reduce the amount of data loss seen by the system. |
Fix |
34083 |
0.1.26 |
Improved rare issue that prevented the CLC from being able to boot up properly after upgrading CLC firmware. |
Enhancement |
34284 |
0.1.26 |
Fixed issue where CLC diagnostics page would list multiple System IP Addresses. |
Fix |
32955 |
0.1.25 |
Fixed an issue where data loss would occur or an implant could drop off the system while in range of a Transceiver (TRX). |
Fix |
23255, 29652 |
0.1.25 |
Improved network performance between the CLC and the acquisition system to prevent disconnects. |
Enhancement |
30172 |
0.1.25 |
Improved system reconnection following a power loss to the CLC. |
Enhancement |
30169 |
0.1.25 |
Improved 3rd part software communication by allowing users to enter IP address for the Network Time Protocol (NTP) server. |
Enhancement |
28080 |
0.1.25 |
Prevention of unnecessary system warnings and messages from being displayed to the end user. |
Fix |
16618 |
0.1.25 |
Support added for PhysioTel Digital product release in other geographies. |
New Feature |
25452 |
0.1.25 |
Corrected the implant voltage value displayed in the PhysioTel Digital configuration dialog. |
Fix |
23187 |
0.1.25 |
Enhancements were made to the system radio frequency (RF) performance between the implant and TRX to reduce data loss. |
Enhancement |
665 |
0.1.24 |
Red LED would illuminate if the CLC failed to obtain an IP Address upon boot. |
Enhancement |
648 |
0.1.24 |
Improved the speed at which the CLC system log is downloaded. |
Enhancement |
16931 |
0.1.24 |
Added the Clear Channel Assessment checkbox to the PhysioTel Digital Diagnostics page. |
Enhancement |
664 |
0.1.24 |
Fix for TRXs being listed within the TRX Status within the PhysioTel Digital Diagnostics page after being unplugged from the CLC. |
Fix |
647 |
0.1.24 |
Updated the IP Address used to obtain the CLC system log file. |
Fix |
637 |
0.1.24 |
Fix for TRX Status information within the PhysioTel Digitial Diagnostics displaying blank information after the TRX is reset. |
Fix |
18800 |
0.1.24 |
Improved CLC communication with NTP server. |
Fix |
18156 |
0.1.24 |
Improved CLC communication with implant. |
Fix |
17810 |
0.1.21 |
Fix for TRX entering an unusable state upon updating the TRX firmware. |
Fix |