MX2 Firmware Release Notes
Model: Matrix 2.0 (MX2)
Version: 2.1.164
Date: June 2019
Firmware Release Notes for MX2 v2.1.164 indicate revisions made to the MX2 Firmware version since the release of the MX2.
Compatibility
MX2s running firmware v1.12.190 are compatible with all PhysioTel Legacy and PhysioTel HD implants, and Ponemah versions 5.20-SP5+ and 6.00+.
Release History
Version |
1.0.7* |
1.0.8 |
1.0.9 |
1.0.10 |
1.0.11 |
1.12.190 |
2.0.97 |
2.0.122 |
2.1.164 |
Release Date |
Dec |
DSI Internal Release |
Feb |
June 2016 |
Jan |
Sept |
June |
June |
Aug |
*Original firmware released with MX2.
New features, enhancements, and fixes
Reference # |
Release Version |
Description |
Disposition |
45348 |
2.1.164 |
Fix added for implants not transitioning from their Pending state to Acquiring upon starting acquisition. |
Fix |
44899 |
2.0.122 |
Fix added for potential false double-pulse near top of second when using digital input samples. |
Fix |
42581 |
2.0.97 |
Added support for the DSI Signal Interface. |
Enhancement |
43916 |
2.0.97 |
Added Finished goods number and revision the Diagnostic homepage. |
Enhancement |
42351 |
1.12.190 |
The NTP Configuration file was updated reduce the polling time from 64 seconds to 8 seconds to more quickly react to time synchronization changes. |
Enhancement |
42421 |
1.12.190 |
Updated the MX2 host name from a generic name to the device name to permit log files to be more easily read. |
Enhancement |
39284 |
1.12.190 |
Updated the firmware change version scheme to include a unquie build number; i.e. changing from 1.0.11 to 1.12.x, where ‘x’ is the build number. |
Enhancement |
39282 |
1.12.190 |
An system verification step was added to the firmware upgrade process to ensure the update is written correctly. If a failure occurs, the system will automatically retry. |
Enhancement |
37539 |
1.12.190 |
Fix was added for a mismatch in error codes which caused a misinterpretation of the message being sent from the MX2 to the system. |
Fix |
37184 |
1.0.11 |
Fix for the MX2 getting into a state where it would never connect to the network or lose communication with the network after a few minutes. When the MX2 was in this state, the user would see the Power, Status, and Error LEDs illuminated on the front panel. |
Fix |
37021 |
1.0.11 |
A button has been added to the MX2 diagnostic webpage’s System Status page to allow the user to more easily download the MX2 log files. |
Enhancement |
34407 |
1.0.10 |
If Ponemah was actively sampling from an implant that was in OFF mode or not within the reception range of the receiver for an extended period of time, additional implants on the acquisition system would lose data. This has been corrected. |
Fix |
34286 |
1.0.10 |
Fixed issue where MX2 diagnostics page would list multiple System IP Addresses. |
Fix |
34084 |
1.0.10 |
Fixed rare issue that prevented the MX2 from being able to boot up properly after upgrading MX2 firmware. |
Fix |
32902/32500 |
1.0.9 |
During data acquisition, communication processes internal to the MX2 hardware may have led to sporadic data dropout. Dropout caused by this issue would impact <1% of the total data collected. This has been corrected. |
Fix |
28697 |
1.0.9 |
Intermittent spikes in the Signal Strength signal during data acquisition has been fixed. |
Fix |
27819 |
1.0.9 |
The Signal Strength (RSSI) webpages have been updated to appropriately format the graph labels when using any web browsers. |
Fix |
27629 |
1.0.9 |
The description of the Matrix 2.0 has been updated to MX2 within the MX2 Diagnostics page device selection dropdown. |
Fix |
30814 |
1.0.9 |
The MX2 clock synchronizes to the acquisition computer once it has been added to the MX2 Configuration. Once synchronized, the MX2 would intermittently not be found by the acquisition computer. This issue was time dependent, causing the intermittent nature of the issue. This has been corrected. |
Fix |
30725 |
1.0.9 |
Added the ability to manually configure the NTP Server's IP address within the MX2 Diagnostics Network page. |
Enhancement |
Comments
0 comments
Please sign in to leave a comment.