QRS has smaller amplitude than than the T wave and Ponemah in incorrectly marking T waves as R waves.
Issue: Ponemah is placing R markings on the T wave due to my QRS complex having smaller amplitudes.
SOLUTION:
Enable the hidden analysis attributes that are not normally displayed. These are normally hidden because they are very rarely required and add un-needed complexity. Please refer to the How to enable ECG 5.3x and 6.x hidden attributes and graph presentation signals article to enable the hidden settings. Also, please refer to ECG Attribute Analysis Description article for more information about the ECG analysis attribute settings.
In the Hidden2 tab, decrease the R width max. Please note that this may affect detection of ventricular ectopic beats.
Can't find what you're looking for? Contact Us