How to enable ECG 5.3x and 6.x hidden attributes and graph presentation signals
Usage
IMPORTANT NOTE
The use of Hidden Presentation signals should NOT be needed and is expected to be rare, but may be necessary in some challenging data sets. The smoothed and absolute derivative presentations are available in the Analysis Attribute dialog box during Review.
Enabling Hidden Attributes
To enable the ECG hidden attributes:
- Navigate to the Ponemah runtime directory.
Default: C:\Ponemah\ - Open the PPP3.INI file.
- Enter the following under the [ECG] heading:
ShowEcgHiddenAtbsFlag=1 - Save the file and exit.
NOTE: Please refer to ECG Attribute Analysis Description for information /description on the ECG attribute settings (both visible and hidden).
Enabling Hidden Presentation Signals
In 5.x the presentation signals Smooth ECG (ECGs), Smooth ECG derivative (ECGsd), and Absolute derivative (AbsDeriv) are always available. In 6.x the DebugMode flag can be used to enable these signals using the process below.
- Navigate to the Ponemah runtime directory.
Default: C:\Ponemah\ - Open the PPP3.INI file.
- Enter the following under the [ECG] heading:
DebugMode=1 - Save the file and exit.