Care and Cleaning the Nebulizer Head
Proper cleaning and maintenance of the nebulizer head will help ensure reliable performance and extend the life of the device.
Routine Cleaning
Clean the nebulizer head after each use.
- Remove the nebulizer head from the plethysmograph or aerosol block.
- Rinse the head thoroughly in warm, soapy water.
- Allow the device to soak for several minutes if needed.
- The nebulizer head may be fully submerged during cleaning.
- Rinse thoroughly to remove any soap residue.
Important: Do not use a high-pressure water stream. Allow water to flow gently through the device during rinsing.
Decontamination
Decontamination procedures depend on the type of nebulizer head being used.
Aerogen® Lab Nebulizer Head
If an Aerogen Lab Nebulizer Head has been used with pathogenic materials and requires decontamination:
- Disconnect the nebulizer head from the power cord.
- Autoclave the unit using a flash steam sterilization cycle:
- Temperature: 134°C
- Duration: 4 minutes
Aerogen® Solo Nebulizer Head
The Aerogen Solo Nebulizer Head is not autoclavable.
If an Aerogen Solo Nebulizer Head has been exposed to pathogenic materials, it should be disposed of following your organization's biohazardous waste procedures. Do not attempt to sterilize or reuse the device after pathogen exposure.
Cleaning Precautions
⚠️ Do Not:
- Use alcohol or alcohol-based cleaning solutions. These can damage the metal membrane.
- Touch the metal membrane.
- Brush, scrub, or apply pressure to the membrane surface.
- Use high-pressure water jets or sprays.
✅ Do:
- Use warm, soapy water for routine cleaning.
- Rinse the membrane gently from both the top and bottom surfaces.
Drying the Nebulizer Head
After cleaning:
- Shake off excess water.
- Gently blot dry with a paper towel or Kimwipes®.
- Place the nebulizer head on its side in a clean, dry location.
Correct Drying Position |
Incorrect Drying Position |
|
Place the nebulizer head on its side.
|
Do not:
Residual moisture can become trapped inside the device, potentially causing discoloration of the metal membrane.
|